*** MIDAS logfile (vers. 02FEBpl1.0) *** Sa, 26 Oct 2002 17:21:31 Page 1 *** MIDAS logfile (vers. 02FEBpl1.0) *** Sa, 26 Oct 2002 17:22:07 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#31 TAKEN ON 021026 IN fastcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 608.38 stddev = 3.23 FIERA temperature = 273.15 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_fastcd_31_021026 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 4.21 1664.4 1665.3 608.5 0.02 0.01 2.74 2 8.41 3271.2 3272.1 1178.5 0.02 0.02 2.78 3 16.81 6479.3 6481.1 2345.9 -0.06 -0.04 2.76 4 25.21 9686.6 9686.1 3489.2 -0.10 -0.10 2.78 5 37.82 14504.6 14505.0 5256.9 -0.10 -0.09 2.76 6 50.42 19329.8 19328.5 7141.3 -0.04 -0.04 2.71 7 63.03 24162.6 24159.1 8853.0 0.01 0.00 2.73 8 84.04 32219.2 32215.7 11693.1 0.07 0.07 2.76 9 105.05 40270.0 40265.1 14375.3 0.09 0.09 2.80 10 126.05 48301.0 48301.8 17149.7 0.08 0.09 2.82 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 4670 to 135488 e- Average count rate measured: 382 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.81 +/- 0.02 e-/ADU RON = 9.07 +/- 0.06 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.142 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.09/-0.10 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay