*** MIDAS logfile (vers. 02FEBpl1.0) *** Fr, 16 Aug 2002 15:50:57 Page 1 *** MIDAS logfile (vers. 02FEBpl1.0) *** Fr, 16 Aug 2002 15:51:40 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#31 TAKEN ON 020816 IN fastcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 609.04 stddev = 3.33 FIERA temperature = 273.15 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_fastcd_31_020816 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 4.21 1712.1 1712.8 622.1 0.07 0.06 2.75 2 8.41 3360.6 3362.0 1242.9 -0.06 -0.04 2.70 3 16.81 6662.7 6662.7 2418.5 -0.05 -0.06 2.75 4 25.21 9960.6 9960.0 3655.5 -0.10 -0.10 2.72 5 37.82 14917.0 14916.5 5537.0 -0.08 -0.08 2.69 6 50.42 19876.7 19875.9 7232.2 -0.04 -0.04 2.75 7 63.03 24848.4 24849.2 9062.8 0.02 0.02 2.74 8 84.04 33130.7 33129.5 12034.9 0.07 0.07 2.75 9 105.05 41415.1 41409.9 15027.8 0.11 0.10 2.76 10 126.05 49668.0 49665.3 17792.4 0.08 0.08 2.79 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 4762 to 138101 e- Average count rate measured: 393 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.78 +/- 0.01 e-/ADU RON = 9.25 +/- 0.04 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.141 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.10/-0.10 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay