*** MIDAS logfile (vers. 01FEBpl1.1) *** Th, 07 Feb 2002 16:37:35 Page 1 *** MIDAS logfile (vers. 01FEBpl1.1) *** Th, 07 Feb 2002 16:38:12 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#31 TAKEN ON 020207 IN fastcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 609.11 stddev = 3.24 There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_fastcd_31_020207 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 4.21 1780.2 1782.3 666.8 0.04 0.05 2.67 2 8.41 3496.6 3497.0 1278.4 0.02 -0.02 2.74 3 16.81 6924.1 6924.1 2494.9 -0.08 -0.09 2.78 4 25.21 10352.4 10349.9 3702.9 -0.10 -0.13 2.80 5 37.82 15503.2 15502.1 5668.2 -0.08 -0.09 2.74 6 50.42 20659.5 20660.7 7404.7 -0.03 -0.02 2.79 7 63.03 25823.0 25824.4 9345.4 0.01 0.03 2.76 8 84.04 34431.4 34431.3 12179.3 0.07 0.08 2.83 9 105.05 43028.4 43027.4 15249.1 0.08 0.09 2.82 10 126.05 51614.8 51616.1 18562.7 0.07 0.09 2.78 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 4998 to 144824 e- Average count rate measured: 409 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.81 +/- 0.01 e-/ADU RON = 9.10 +/- 0.05 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.146 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.08/-0.11 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay