EFOSC2 

ESO Faint Object Spectrograph and Camera
 
 
  
Spectroscopic Science Modes

Long slit spectroscopy
Long slit spectroscopy - scanning SpectroPolarimetry
Slitless Spectroscopy

This instrument was built primarily for spectroscopy. EFOSC2@3.6m is one of the most efficient instruments available when low resolution (dispersion < 1000) is the need of the hour. The different science spectroscopic modes modes are described below. The observer is also referred to the pages on Acquisition images and Miscellaneous information.

The seeing at the 3.6m/EFOSC2 has never been better than 0.65  - the seeing is more typically between 0.9 and 1.2  -  and so the binned pixel size of 0.314" adequately samples the resolution afforded by all but the 0.5" slit. Thus unless spectral resolution is of first importance and one has bright targets using the CCD in the binned mode is strongly recommended.

Long Slit Spectroscopy

In this mode one can obtain one or more spectra of 1-2 objects through 1 or more grisms. However, it is essential that all templates in an OB have the same slits. There is no provision for moving the object along the slit from one exposure to another.

The following acquisition templates may be used in this mode

A sample jP2PP observing block

The above example shows an OB comprising a sequence of:

The on-source internal flats may be needed for Grism #12 (and other red grisms) to correct for the strong fringing
seen in their spectra. While CCD flexure with pointing is in general much smaller than a pixel, it may still be sufficient to result in fringe residuals after flat-fielding. An on-source Helium-Argon arc lamp spectrum may not be needed unless the science goal is to get accurate radial velocities since the flexure is much less than a pixel.

The FITS file is called  EFOSC_Spectrum.#.fits

Parameter Values   One does not have to (in fact, should not) change the value of any parameter other than:

Long Slit Spectroscopy - Scanning

This is very similar to Long Slit Spectroscopy except that the telescope is moved across the extended target while integrating in order to sample the spectrum across a larger part of the target. There are 2 ways in which this can be achieved:

1. Unguided scanning across the target
One can apply a velocity to the telescope in the RA and/or DEC directions after the shutter has opened and with the AutoGuider turned off. This is to be used when the telescope scans many tens or hundreds of acrseconds during an exposure or if the required velocity is large (say, less than 0.1 arcsec/sec). Since the telescope is not being guided the motion is not precise and one may have a 1" wander around the desired track.The problem with large translations is that the guide star may fall off the edge of the guide probe field. In the case of large velocities the star may move out of the target square between one guide correction and the next (done every 5 seconds). Follow the recipe given below :

One may have any number of templates in the OB but each template should ask for just 1 exposure because the OB has to be paused and the telescope brought back to the starting position before starting the next exposure. The grisms used may (or may not) change from one template to another. If different exposure times are used the telescope velocity has to be modified accordingly.

You will see the guide star starting to move away from the guide target area once the guiding is switched off. If one wants to repeat the exposure,  with the same grism or another one, one can ask the telescope operator to offset the telescope so as to bring the guide star back into the guide target area - this will ensure that the telescope is again pointing to the starting position. This should be much faster than requiring the image with an acquistion template. An alternative is to repeat the telescope movement backwards for the next exposure.

2. Differential tracking

One can apply a constant velocity to the guide probe relative to the telescope. Since the AutoGuider is turned on this moves the telescope in the opposite direction in a controlled manner. The recipe is similar to that for the first case except for

  • Keep the guiding turned on
  • Type the required velocity value on the AutoGuider panel (under differential tracking).
  • Start the  movement of the guide probe
  • Some other Issues

    Try and reduce the scan speed so as to be able to use the differential tracking method; aligning the slit along the major axis of the extended source is one way of soing so.

    Important   The OBs have a box for differential RA and DEC tracking (on the bottom right of the panel). These values are not considered while executing the templates. The telescope movement has to be started manually (by the operator)

    Always confirm the direction of motion of the telescope - there are several sources of confusion : whether the velocity refers to the movement of the telescope or the movement of the source across the field (in the first case) and whether it refers to the movement of the  telescope or that of the guide probe (in the second case). One should ask the support astronomer to confirm the sense by trailing a star on the image with a velocity of known sign.

    These observations require constant communication between the observer and the telescope operator and manual intervention by the latter over and above the regular actions - there is scope for a lot of confusion. Try to keep the intervention as simple as possible by, for example, confining the scan to just RA or DEC and providing clear instructions to the telescope operator.

    SpectroPolarimetry

    The SpectroPolarimicetr mode is very similar to the Long Slit Spectroscopy  mode  except for the following differences:

    Refer to the page on Polarimetric optical elements as well.

    Any one of the 2 available SpectroPolarimetric Masks/Slits  may be mounted on the aperture wheel but the name in jP2PP will be the same. The Wollaston Prism in the grism wheel must be aligned in a manner appropriate for a parallel mask.

    Important  When the half wave plate (HWP) is set to continuous rotation in a template the counter ends up with 360xN degrees. When a signal is sent to set the HWP to some small fixed value (say, 45 degrees)  it tries to undo all the N rotations and in doing so triggers a time-out error. So we have included a command at the end of every Polarimetry template to initialise the HWP if it was set into continuous rotation mode - this just takes a few seconds. However when such a template is aborted, for whatever reason, this initialisation does not take place. The only solution is to cold-start the EFOSC2 instrument which takes about 1-2 minutes. So avoid aborting a Polarimetry template - use the STOP option instead.

    A sample jP2PP observing block

    The above example shows an OB comprising a sequence of

    The FITS file is called EFOSC_SpecPol.#.fits

    Parameter Values   One does not have to (in fact, should not) change the value of any parameter other than:

    Slitless Spectroscopy

    The use of filters in combination with a grism can be useful to isolate the spectral region of interest and to reduce crowding and sky background intensity. Note that the spectral coverage depends on the position of the objects. Measures are needed to define the wavelength ranges as a function of the object position and the overall efficiency in this mode.
    back to top


    Send comments to : ls-spectro

    Last modified: Mon Jan 14 21:29:03 CLST 2002 

    [3.60 & CAT Homepage][La Silla][ESO][Observing][Search][Help][News]