![]() |
EFOSC2ESO Faint Object Spectrograph and Camera |
---|
Acquistion Images |
|
It is usually the case that one wants to locate the target, in both imaging and spectroscopy, on a specific pixel of the CCD. This is achieved by executing an Acquisition Template. One acquisition template (and only one) is an essential part of a science observing block - it is always the first template on a science OB and may be followed by one or more science templates.
More details on acquisition images may be found for each science observing mode on the corresponding web page (Imaging and Spectroscopy).
An actions executed by an acquisition template can be divided into 3 well-defined parts.
1. one can ask the telescope operator to skip the acquisition template - in this case the science templates are executed immediately, orTelescope Focus The telescope must be focussed every time it is moved by 15-20 degrees on the sky - either because the target was changed or because the telescope moved substantially while tracking a target. Of course any change in temperature will necessitate a refocus. It is not necessary to refocus from one filter to another.
2. one can set the Preset flag to False - in this case the telescope skips to the Focus part in the above sequence. Important - if you want the adaptor angle changed the Preset flag must be set to True even if you are on target
Acquisition Exposure Time It takes between 1 and 3 iterations to get the positioning of the target exactly right - and it is a strong function of how many photons the object has on the image. Always err on the side of caution by specifying a larger acquition exposure time than you think is necessary - it takes less time to keep the shutter open for another 20-40 seconds than to go through another iteration! The default 20 second exposure is sufficient for objects up to 19-20 mag (in R). Scale up the exposure correspondingly for fainter objects.
The minimum exposure time allowed by the template is 1 second. However ultra bright stars (e.g. spectrophotometric standards) which will saturate the CCD even with this exposure and make positioning on the slit difficult. One solution would be to request the telescope operator to make the change (down to 0.1 second) directly on BOB.
Using Acquisition Images for Photometry The acquisition images are always read out in the fast mode to save time. This means that such frames have a different gain, bias and readout noise than frames taken in the normal readout mode. We recommend that observers not use these images for science unless they have calibration frames (bias, flat fields etc) for the fast mode. Also note that the fast mode is read using 2 different amplifiers which means that the two halfs of the CCD have different biases and gains.
Sometimes observers specify long acquisition exposure times with the
intention of doing photometry on the fainter objects in the field. Remember
that if you go through several iterations (for whatever reason) to position
your target you will be wasting time because every one of those exposures
will use that long exposure time. It is better to take short (but sufficient)
acquisition and then add a regular science observing template to it.
|
jP2PP panel with typical parameters
The above example shows a simple preset template which makes the telescope
track the target with the y-axis of the CCD aligned 25 degrees west of
north.
1. widely scattered objects (eg. a large cluster, standard star cluster) are all are positioned on the CCDThe object positioning is done as follows :
2. essential objects do not fall on bad pixels
jP2PP panel with typical parameters
The above example shows an acquisition template which does a Preset (points the telescope to the target with the adaptor angle = 270 + 0 = 270 degrees and starts tracking) followed by a Focus procedure and finally moves the desired object to pixel (1100, 1024)
The FITS image is named EFOSC_AcqPix.#.fits
jP2PP panel with typical parameters
The above example shows an acquisition template which does a Preset (points the telescope to the target with the adaptor angle = 270 + 0 = 270 degrees and starts tracking) and without doing a focus goes on to moving the desired object to a pixel which has to be defined when the procedure is run. The acquisition is done with the Half Wave Plate in position.
The FITS image is named EFOSC_AcqPol.#.fits
|
Note that even though the acquisition frame is an image, taken without a slit, one still has to specify the slit intended for use. This is because the y-pixel location of different slits are different (the y location is recorded in a database when a slit is mounted in the instrument). Additionally, one also has to specify the x-pixel along the slit where the target will be positioned. One should chose a column which has no bad pixels as the x-pixel value.
The x-pixel value for spectropolarimetry should be determined by following the procedure described under Identifying the Acquisition reference pixel
One can either orient the slit in a particular direction (along the major axis of a particulat source) or let the procedure align it along the parallactic angle. One selects for the parallactic angle by choosing -9999 for the rotator offset angle in the template. The other option is to input an rotator offset angle (ROA) which has to be between -100 and +80 degrees. The position angle of the slit is given by PA = 90 + ROA. Important: it is advisable to sellect the parallactic angle when the object is low in the sky, especially when conducting spectroscopy with a large wavelength range, so to avoid losing flux due to atmospheric refraction. The importance of atmospheric differential refaction in spectrophotometry is discussed by Filippenko (1982).
The object positioning is done as follows :
jP2PP panel with typical parameters
The above example shows an acquisition template which does a Preset (points the telescope to the target with the adaptor angle = parallactic angle and starts tracking) followed by a Focus procedure and finally moves the desired object to pixel (1100, y-pixel of slit#1.5)
The FITS image is named EFOSC_AcqSlit.#.fits
Important Since every slit
has a different y-pixel position one cannot mix observations with different
slits in the same OB. One can however mix different grisms but with the
same slit
Note that even though the acquisition frame is an image, taken without a slit, one still has to specify the slit intended for use. This is because the y-pixel location of different slits are different (the y location is recorded in a database when a slit is mounted in the instrument).
The procedure itself calculates the appropriate position angle of the slit that includes both the required objects (see the procedure described below). So the easiest option is to define the rotator offset angle (ROA) = 0 in the template. However observer can save 2 minutes of acquisition time if he/she specifies an approximate rotator offset angle such that the objects are parallel to the slit, especially if the rotator needs to be moved a long way from its default position. However, note that one should be very clear about the adaptor limits and ESO adaptor angle definition before trying this especially when wants to have the slit close to the north-south direction.
Additionally, one also has to specify the x-pixel along the slit where the one of the targets will be positioned. One should chose a column which has no bad pixels as the x-pixel value. Also make sure that with this x-pixel value the second target will not fall on a column with bad values.
The x-pixel value for spectropolarimetry should be determined by following the procedure described under Identifying the Acquisition reference pixel
The object positioning is done as follows :
Rotation to align the slit with the objects:
jP2PP panel with typical parameters
The above example shows an acquisition template which does a Preset (points the telescope to the target with the adaptor angle = 270 + 0 and starts tracking) followed by a Focus procedure and finally rotates the adaptor to align the slit parallel to the 2 objects and then moves the selected object to pixel (1450, y-pixel of slit#2.0)
Important: Since every slit has a different y-pixel position one cannot mix observations with different slits in the same OB. One can however mix different grisms but with the same slit.
The FITS image is named EFOSC_AcqRotSlit.#.fits
|
Given below is a list of acquisition template parameters which should/may be modified:
Send comments to : ls-spectro
Last modified: Wed Jan 16 03:12:05 CLST 2002 |
![]() ![]() ![]() ![]() ![]() ![]() ![]() |