EUROPEAN SOUTHERN OBSERVATORY
La Silla - Science Operations Department
|
|
|
|
FEROS on
the MPG/ESO-2.20m
|
Internal Flatfield Lamp WarmUp Time
It has been known since the days of FEROS at the ESO-1.52m that the
Flatfield Lamps require some time to warmup before their output flux
stablises, however it has apparently never been systematically
investigated how long is actually required to reach the stable
operation regime. This report gives a quick answer to this perenial
question.
Data acquisition and reduction
On the night of 2005-10-03/04, which was otherwise lost to poor
weather conditions, several hours after a standard DRS initialisation
(to allow the FF lamps to fully cool back to room temperature), a
sequence of 30 FFs were acquired in the fast readout mode. These data
were reduced using the FEROS-DRS package available from here using the write/key
RedAllIms/c/1/1 "Y" option to force the DRS to process the
flatfield images.
Results
The plots below clearly show that the FF flux has stabilised
approximately 6mins after the OB is started. Therefore the currently
adopted 5min warmup used in the StanCal OBs and in CalOB is probably
just adequate and reasonable since there is generally of the order of 1
additional minute of overhead due to moving functions etc.
Warning
Presumably the required warmup time can vary according to the
state of the lamp, after many hours of use the required warmup may
increase, therefore this test should be repeated from time to time.