During a warm up of the WFI Dewar during 19 August 2005 the system that controlled the temperature malfunctioned and the Dewar was brought to temperatures in excess of 90 degrees celsius. This contaminated the Dewar window and the CCDs. The system was later decontaminated but it started slow process of re-contamination. Data taken before 19 August 2005 should not be calibrated with data taken after this date, and vice versa. This page was born to keep track of the contamination and to determine the duration/validity of dome and sky flats. As of 20 October 2005 we have seen a change in beta-light frame characteristics of 1.5%/month.
For those of you who would like some additional information, please note the following:
The plot below shows the evolution of the dome flat fields. Each point marks the value of the ratio of the dome flats over a fiducial one. Sixteen areas are ploted: the four corners where the contamination effects are maximal, four areas around the center of the mosaic; and 8 regions one near the centers of each of the chips.
The plot below shows the evolution of the SKY FLAT field frames. Each point marks the value of the ratio of the dome flats over a fiducial one. Sixteen areas are ploted: the four corners where the contamination effects are maximal, four areas around the center of the mosaic; and 8 regions one near the centers of each of the chips.
The spacing between the contours below is 0.1%
|