Logo

CCD38 Linearity Test

Logo

EUROPEAN SOUTHERN OBSERVATORY

CCD LINEARITY TEST for BOLLER & CHIEVENS (CCD #38)

 

 

This Linearity test has to be performed once a week.

At the dome B&CH spectrograph:

On the control room:

At the HP-100 terminal type:

On the Ins1p5 PC exit the current session and begin a new one with:

User : ccd

Pass :

On the Midas session hit enter.

On the new hpterm window type the following command:

ccd_standardtest 38 s (the #38 corresponds to the CCD installed)

Then hit enter…

This will take an exposure, once this is finished follow the instructions on the hpterm window.

On the User Interface, clic on UTILS, then clic over STATISTICS OF REGION (a rectangular area will pop up), then pick an area in the image, select with the left hand button of the mouse, then clic on the mouse middle buttom to kill it. A process will be performed, look for the MEAN STANDARD DEVIATION: e.g. 9.5364647e+10 = ~9531. This value has to be written at the hpterm window prompt, then hit return.

25 images will be taken… (this will take about 30 minutes)

To make sure that all images (*,bdf, *,mt) have arrived , open a new hpterm window and check it out with "ls –1 *.bdf, ls –1 *.mt.

NOTE: If the *,bdf images did not arrive do the following on a hpterm window.

Indisk/fits bch xxxx.mt bchxxxx.bdf (this will tranfer the bdf missing file to the current directory.

You will have to this for every missing file.

When the set of 25 exposures is finished, terminate the current Midas session, type "bye".

Now begin a new midas session, use the following instructions..

Now run the following batch:

@@ ccd_starndard_test 38 010620 bch s xxxx (ß Last image)

Once the test has finished there will be a print out, check out the following parameters in the hard copy:

C.fact = 1.58+/- 0.01e-/adu

Ron = 7.61 +/- 0.04e- rms

These values are approximately the right ones, compare them with the ones posted on the web:

http://www.ls.eso.org/lasilla/sciops/2p2/CCDs/38/index.html

If there is too much difference between your data and the one on the web page, run the test again (make sure everything is fine), if the difference persists, call Detectors Team (4111).

If the data is within the normal values proceed to the WEB UPDATE, send data to "meli"

On a hpterm window type: to_meli 38 yymmdd password, then hit enter… a transfer will occur…

Telnet meli:

User: t2p2web

Passw:

Then at the prompt: meli > update 38

Now go backwards (go to the dome):

In the control room at the HP1000.

NOTE: On the HP1000, on the upper right corner under the plastic protective cover you’ll find 8 keys, use key f8, hit this key until a ccd option appears (f2), once this is done a new display will appear hit key f3 to Initialize ccd, if this is o.k.

At the prompt type:

Now re-start the session: ccd, La Silla

Comments to: atorrejon@eso.org