| Standard Test for CCD 02 (DFOSC)
|
|
Taking the Data
- Install the beta light in the FREE POSITION of the filter
wheel.
- Verify that the main mirror and the sky baffle are closed
and switch off the light in the dome.
- In DAISY, load the sequense standartestsec1. The first 5
images are biases, in this case the instrument setup is not
important, for the next images We are using the setup 2, set the
instrument setup 2 with FILTER WHEEL FREE and GRISM WHEEL FREE
(the aperture wheel position is not important). Then start the
sequence.
- At the end of the previous sequence load the sequence
standartestsec2 and then start the sequence.
- At the end of the standartestsec2 remove the beta light.
NOTE: The standartestsec1 and 2 are in the /data/daisy
directory but in any case there is another copy in the /data/daisy/bin
directory.
Data Reduction
Midas batch (must be MIDAS 95 or later version, on ins1d5 you can
use the command: inmidas -r 95NOV in xterm window) executed using the command:
@@ ccd_standard_test 02 010304 dfsc fast 0666
Where
- ARG1 is the CCD number (02)
- ARG2 is the date (YYMMDD)(010304)
- ARG3 is the image filename prefix(dfsc)
- ARG4 is the readout mode. This should always be `fast' for CCD02.
- ARG5 is the file number of the last image.(0666)
Once the reduction is complete, send the data to meli using:
to_meli 02 YYMMDD meli_password
Then, log into meli and update the pages with the command:
update 02