CCD Tests

Overview

A complete set of CCD tests can be obtained, processed and the relevant web pages updated by typing just two commands, one in BIAS and one in MIDAS.

Procedure

  1. At FEROS, move the Beta Light into position in front of the FEROS Camera.
  2. Login on on the BIAS-PC: User "feros", Passwd <see list>.
    Once BIAS is up and running you will see four windows; BIAS Obs DataBase panel, BIAS CCD CONTROL, BIAS CCD STATUS, and SAOImage bias_ds9, plus the  BIAS CCD LOG window will be iconised). In the BIAS Obs DataBase panel set the Program Id. == "00.A-0000(A)", PI == "Test", Observer == "<Your Initials>".
    BIASPC-BIAS
    Note: Only ONE instance each of BIAS and the fcu panel can be running at any time. If you try to start BIAS while another instance is already running it will fail with NO warning. If you try to start the fcu panel while another instance is already running the panel will appear but it will be useless and there will be many error messages during startup, simply close the panel.
  3. Run the test script(s):
  4. Open a NightLog window which will be useful to monitor progress and eventually check on the file numbers of the images:
  5. Open a CCDTest MIDAS session:
  6. Run the CCDTest analysis program. In the MIDAS window type:
    where:
    After reducing all the tests, you will be prompted for the password for the t2p2web account on Meli so that the results can transfered. The updated Web Page will be generated for each Amplifier/Gain Combination.
    Note: Do Not change the desktop while MIDAS is reducing the test data as this will cause problems with the generation of PS files of the images.
    Note: If there were no problems with the data acquisition, i.e. the 150 images are named with the 150 sequential numbers then just the number of the first test image followed by 5 -1s can be used to specify the numbers, e.g.
    1,26,51,76,101,126 could also be written as 1,-1,-1,-1,-1,-1.
  7. Remove the Beta Light and replace the cap.

TroubleShooting

A few problems are known to occur from time to time during the execution of the CCDTests. Here are descriptions and solutions (if you come across another problem please email j.pritchard@eso.org with the details)
  1. BIAS hangs during image acquisition
    This problem results when the CCD electronics completes an exposure faster than the BIAS software can keep up with, e.g. when the CCD window is particularly small and a short exposure time, e.g. a 10x10 pixels bias image.
    Symptom: BIAS CCD STATUS is "Pending"
    Cure: The only known cure is to "kill" BIAS. This can be done simply by exiting from DS9.
    Note: This problem often happened during CCDTests. Hopefully with the reverting of the BIAS PC back to the 400MHz computer from the 200MHz one, it will happen less often. The network can also play a big part in this problem, so the best place to run the CCDTests from is the BIAS-PC itself, rather than w2p2oh, or even worse at the RITZ.
    Note: If this problem does occur, please restart BIAS and then continue with the CCDTests. If it fails during the first amp/gain combination then rerun the test4ccd-all script, but otherwise just rerun the
    amp/gain combinations for which a complete set of data was not yet completed, remember the sequence of amp/gain combinations is; blow, alow, ablow, bhigh, ahigh, abhigh. In this case check the log file (the file FEROS.<YYYY-MM-DD>.log in the log subdirectory of the feros user's home directory on the BIAS-PC) carefully to ensure you get the correct first image number for each amp/gain combination.
  2. Failure of the Analysis
    Usually this problem results from either entering the wrong number for the first image of the test, or because there is an incomplete data set for a give amp/gain combination.
    Symptom
    : MIDAS reduction and analysis fails.
    Cure: Check carefully the log file (the file FEROS.<YYYY-MM-DD>.log in the log subdirectory of the feros user's home directory on the BIAS-PC) to ensure that you have the correct first image number for each amp/gain combination and that each amp/gain combination has 5 bias images and 20 FlatField images.

Description of MIDAS Procedures

ferosProcessAllCCDTests

  1. Usage:
    where
  2. Overview:
  3. TBD.

ferosProcessCCDTestImages

  1. Usage:
    where
  2. Overview:
  3. TBD.