*** MIDAS logfile (vers. 97NOV) *** We, 29 Dec 1999 17:34:20 Page 1 *** MIDAS logfile (vers. 97NOV) *** We, 29 Dec 1999 17:35:31 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#60 TAKEN ON 991229 IN blow MODE ========================================================== BIAS ==== 5 biases at start test: average = 200.55 stddev = 2.12 There were 1 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_blow_60_991229 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 2.00 756.7 757.1 304.4 0.22 0.22 2.49 2 4.00 1495.4 1495.5 600.0 -0.39 -0.40 2.49 3 8.00 2989.4 2989.4 1204.6 -0.13 -0.14 2.48 4 16.00 5972.7 5972.3 2378.3 -0.09 -0.10 2.51 5 32.00 11953.3 11953.5 4670.2 0.05 0.06 2.56 6 52.00 19440.7 19441.0 7478.0 0.17 0.17 2.60 7 72.00 26909.2 26910.1 10191.2 0.15 0.16 2.64 8 92.00 34373.0 34374.4 12661.8 0.12 0.13 2.71 9 112.00 41805.1 41803.5 14995.4 0.03 0.03 2.79 10 130.00 48441.8 48440.8 16746.8 -0.13 -0.13 2.89 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 2151 to 137678 e- Average count rate measured: 373 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.84 +/- 0.05 e-/ADU RON = 6.04 +/- 0.11 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.024 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.22/-0.39 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay