*** MIDAS logfile (vers. 97NOV) *** Fr, 08 Mar 2002 18:06:21 Page 1 *** MIDAS logfile (vers. 97NOV) *** Fr, 08 Mar 2002 18:07:45 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#60 TAKEN ON 020308 IN blow MODE ========================================================== BIAS ==== 5 biases at start test: average = 207.38 stddev = 1.87 There were 1 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_blow_60_020308 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 2.00 626.4 625.4 257.6 0.08 0.07 2.43 2 8.00 2463.7 2463.7 1011.8 -0.23 -0.19 2.43 3 16.00 4918.9 4918.3 1985.7 -0.17 -0.17 2.48 4 32.00 9839.3 9838.3 3946.4 -0.04 -0.05 2.49 5 52.00 16001.2 16001.0 6275.3 0.08 0.08 2.55 6 72.00 22153.8 22153.3 8546.7 0.10 0.09 2.59 7 92.00 28300.9 28303.9 10604.8 0.08 0.09 2.67 8 112.00 34448.5 34449.5 12828.7 0.08 0.08 2.69 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 1685 to 92743 e- Average count rate measured: 307 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.69 +/- 0.03 e-/ADU RON = 5.02 +/- 0.06 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.035 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.09/-0.21 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay