*** MIDAS logfile (vers. 97NOV) *** We, 29 Mar 2000 18:46:43 Page 1 *** MIDAS logfile (vers. 97NOV) *** We, 29 Mar 2000 18:47:54 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#60 TAKEN ON 000329 IN blow MODE ========================================================== BIAS ==== 5 biases at start test: average = 202.43 stddev = 2.12 There were 1 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_blow_60_000329 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 2.00 748.5 748.8 303.0 0.20 0.20 2.47 2 4.00 1479.3 1479.3 588.1 -0.32 -0.34 2.52 3 8.00 2954.9 2953.9 1193.5 -0.11 -0.15 2.48 4 16.00 5900.7 5900.5 2346.6 -0.10 -0.10 2.51 5 32.00 11807.1 11807.9 4633.2 0.03 0.05 2.55 6 52.00 19204.1 19203.2 7408.7 0.16 0.16 2.59 7 72.00 26580.4 26580.2 10070.1 0.13 0.14 2.64 8 92.00 33954.4 33952.8 12625.0 0.11 0.12 2.69 9 112.00 41306.6 41302.0 14738.2 0.05 0.05 2.80 10 130.00 47852.0 47850.5 16698.8 -0.14 -0.14 2.87 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 2118 to 135379 e- Average count rate measured: 369 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.83 +/- 0.05 e-/ADU RON = 5.99 +/- 0.10 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.027 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.20/-0.33 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay