Logo

The FEROS at the MPG/ESO-2.20m CCD Page


CCD#60 in mode Amp A, High gain.

Logo

General Information

The CCD camera is operated with the CCD controller developed by the Astronomical Instrument Centre at the AOC and identical to that in operation at the Danish 1.54-m telescope on La Silla. The controller is capable of delivering the full performance of the CCD in terms of readout noise, CTE, immunity from interference, crosstalk, etc., and provides the image files with the necessary header information. Readout times are of the order of 70 sec through two parallel signal chains.

CCD Characteristics
Op. temperature -1100 to -1200 C
RON 3 e- (for 50 kpix/sec/port)
Dark current 1.5 e-/h/pixel
CTE (par/ser) 0.99999/0.99999
Linearity < 1% deviation up to full well
Full well capacity 150.000 e-

The performance figures given above are based on experience at AOC and are typical measured figures, after optimising operational temperatures, clock waveforms, etc. They apply to CCDs operated at a pixel rate of 50 kpixels/sec and at a nominal temperature of -110°C. ``Full well'' is defined as the exposure limit to which no measurable departure from linearity occurs; the exposure level causing actual blooming is considerably higher.

The camera uses an EEV CCD detector of the design specified for FEROS, i.e., 2048x4096 pixels of 15x15microns, mounted in a continuous-flow cryostat. The CCD can only be mounted with the columns in the main dispersion direction.

The thinned 2kx4k CCD and the standard ESO-VLT continuous-flow cryostat are provided by ESO.

The CCD controller electronics (+ 1 spare) were built by the AOC; also the integration and testing of the complete CCD system were carried out at the AOC.

Note regarding the bias level from Peter Sinclair:

My experience with the Danish controller on DFOSC tells me that a change in 5 adu is rather normal... Actually, the bias level of DFOSC moves around a larger figure... within a single test. I have discussed this with Preben Norregard and he says that this should not be a problem provided the user takes the bias level from the overscan area...

* Derivatives
* Standard CCD Test
* Long CCD Test



Time Evolution

The below plot shows the variation of the bias level, standard light-source (note that often a beta-light is used which, due to radioactive decay, will fade. It is also possible that the phosphor can be poisoned, but our data show no evidence for such an effect), gain, read-out noise, shutter error, and linearity error as a function of time. The values for the bias, gain and read-out noise are plotted along with the associated measurement error. The blue dashed line in the plots shows the mean value of all the points, while the dotted blue line shows the standard deviation of this distribution. The blue and red points in the bottom plot show the maximum deviations in a positive and negative sense wrt the least-squares fit of the linearity data. Plotted vertically on the bottom of the graph is the date (dd/mm) on which the test was made.

[Click here to download ps file]

For printing, the above plot is also available in postscript format . The raw data can also be viewed in a tabular format.




Standard CCD Test

Below we show the most recent test.



*** MIDAS logfile (vers. 02FEBpl1.0) *** Tu, 17 Jun 2003   01:33:26   Page 1
 
 
*** MIDAS logfile (vers. 02FEBpl1.0) *** Tu, 17 Jun 2003   01:33:53   Page 2
 
     ==========================================================
     SUMMARY STANDARD TEST CCD#60 TAKEN ON 20030616 IN AHIGH MODE
     ==========================================================
 
      BIAS
      ====
 
      5 biases at start test: average = 423.1682 stddev =  14.5803
 
      There were 186 hot pixels/columns detected
 
 
      TRANSFER FUNCTION AND LINEARITY DATA
      ====================================
 
  Table : Std_AHIGH_60_20030616 
 Sequence Exp_tim1  Med_cnt1  Med_cnt2  Variance  Frc_cnt1  Frc_cnt2  el_adu1  
 -------- --------- --------- --------- --------- --------- --------- ---------
        1     1.000 1686.4919 1687.6420 3335.3750    0.9971    0.9976      0.25
        2     2.000 3358.3320 3359.1421 6364.2998    0.9977    0.9977      0.26
        3     3.399 5704.3818 5706.0820 10744.099    0.9992    0.9993      0.27
        4     6.099 10232.631 10233.131 19252.400    1.0002    1.0000      0.27
        5     9.500 15929.332 15933.631 29606.900    1.0002    1.0003      0.27
        6    13.699 22977.832 22978.531 41897.148    1.0008    1.0007      0.27
        7    18.600 31201.931 31199.832 57360.000    1.0011    1.0009      0.27
        8    24.300 40763.433 40750.632 74971.500    1.0013    1.0008      0.27
        9    30.800 51651.832 51674.433 94181.000    1.0010    1.0013      0.27
       10    35.000 58717.632 58724.632 107018.00    1.0015    1.0014      0.27
 -------- --------- --------- --------- --------- --------- --------- ---------
 Sequence el_adu2  
 -------- ---------
        1      0.25
        2      0.26
        3      0.27
        4      0.27
        5      0.27
        6      0.27
        7      0.27
        8      0.27
        9      0.27
       10      0.27
 -------- ---------
 
      Dynamic range explored: from 928 to 32290 e-
      Average count rate measured: 1675 ADU/sec.
 
      CONV. FACTOR AND RON
      ====================
 
      C.Fact = 5.49884E-01 +/- 3.83438E-03 e-/ADU
 
      RON    = 8.01746E+00 +/- 5.59063E-02 e- RMS
 
 
      SHUTTER ERROR
      =============
 
 Average shutter error derived from fractional count rate = 1.00000E-02 sec
 
*** MIDAS logfile (vers. 02FEBpl1.0) *** Tu, 17 Jun 2003   01:33:53   Page 3
 
 
 
      LINEARITY ERROR
      ===============
 
 Average linearity error after shutter correct. = + 1.0014/ 0.9973 %
 
 These errors represent maximum deviation in % from the average
 count rate after correction for shutter delay

In addition, the previous standard tests are available. Click in the table below to access these files.

Standard Tests 1 2 3 4 5 6
20030211.txt aver.ps cbias.ps hist.ps lin1.ps lin2.ps trans.ps
20030212.txt aver.ps cbias.ps hist.ps lin1.ps lin2.ps trans.ps
20030303.txt aver.ps cbias.ps hist.ps lin1.ps lin2.ps trans.ps
20030304.txt aver.ps cbias.ps hist.ps lin1.ps lin2.ps trans.ps
20030325.txt aver.ps cbias.ps hist.ps lin1.ps lin2.ps trans.ps
20030326.txt aver.ps cbias.ps hist.ps lin1.ps lin2.ps trans.ps
20030327.txt aver.ps cbias.ps hist.ps lin1.ps lin2.ps trans.ps
20030328.txt aver.ps cbias.ps hist.ps lin1.ps lin2.ps trans.ps
20030329.txt aver.ps cbias.ps hist.ps lin1.ps lin2.ps trans.ps
20030330.txt aver.ps cbias.ps hist.ps lin1.ps lin2.ps trans.ps
20030401.txt aver.ps cbias.ps hist.ps lin1.ps lin2.ps trans.ps
20030402.txt aver.ps cbias.ps hist.ps lin1.ps lin2.ps trans.ps
20030403.txt aver.ps cbias.ps hist.ps lin1.ps lin2.ps trans.ps
20030504.txt aver.ps cbias.ps hist.ps lin1.ps lin2.ps trans.ps
20030505.txt aver.ps cbias.ps hist.ps lin1.ps lin2.ps trans.ps
20030506.txt aver.ps cbias.ps hist.ps lin1.ps lin2.ps trans.ps
20030507.txt aver.ps cbias.ps hist.ps lin1.ps lin2.ps trans.ps
20030601.txt aver.ps cbias.ps hist.ps lin1.ps lin2.ps trans.ps
20030602.txt aver.ps cbias.ps hist.ps lin1.ps lin2.ps trans.ps
20030605.txt aver.ps cbias.ps hist.ps lin1.ps lin2.ps trans.ps
20030606.txt aver.ps cbias.ps hist.ps lin1.ps lin2.ps trans.ps
20030607.txt aver.ps cbias.ps hist.ps lin1.ps lin2.ps trans.ps
20030608.txt aver.ps cbias.ps hist.ps lin1.ps lin2.ps trans.ps
20030615.txt aver.ps cbias.ps hist.ps lin1.ps lin2.ps trans.ps
20030616.txt aver.ps cbias.ps hist.ps lin1.ps lin2.ps trans.ps

In the above table, the postscript files are:

Column
Number
Description
1 Average of the rows and cols. of the bias image
2 Combined bias frame of 5 bias images
3 Histogram of the bias image
4 Plots of exposure sequence, unedited transfer and raw linearity
5 Plots of filtered linearity and differential non-linearity in
6 Filtered transfer curves for both data sets

Note: filtered means that saturated images or deviating too much from the other data, if present, have been removed.




Long CCD Test

Below are the results of the long tests carried out by ODT. Be warned that these are large compressed tar files, and are of specialised interest. To download, press the shift key and click on the hypertext link with your middle mouse button.

Long Tests


Updated:
16/Jun/2003
2p2 Team
CCD Page
2p2 Team
Home
Comments to:
ls-feros@eso.org