*** MIDAS logfile (vers. 02FEBpl1.0) *** Mo, 08 Sep 2003 17:47:24 Page 1 *** MIDAS logfile (vers. 02FEBpl1.0) *** Mo, 08 Sep 2003 17:47:44 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#60 TAKEN ON 20030908 IN BLOW MODE ========================================================== BIAS ==== 5 biases at start test: average = 383.4742 stddev = 1.8521 There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_BLOW_60_20030908 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 1.000 359.3120 359.1040 150.3755 1.0008 1.0002 1.19 2 6.799 2410.7461 2411.3159 984.5950 0.9960 0.9963 1.22 3 15.300 5432.7959 5432.9058 2204.3799 0.9982 0.9983 1.23 4 27.199 9664.1260 9663.7256 3815.0300 0.9992 0.9992 1.27 5 42.500 15115.126 15114.425 6060.9502 1.0002 1.0002 1.25 6 61.200 21782.025 21780.226 8572.8496 1.0011 1.0010 1.27 7 83.299 29659.826 29661.226 11692.250 1.0015 1.0016 1.27 8 108.799 38743.425 38743.324 14584.750 1.0017 1.0017 1.33 9 137.699 49015.226 49016.324 18177.949 1.0013 1.0013 1.35 10 170.000 60442.726 60440.125 21616.150 1.0001 1.0001 1.40 -------- --------- --------- --------- --------- --------- --------- --------- Sequence el_adu2 -------- --------- 1 1.19 2 1.22 3 1.23 4 1.27 5 1.25 6 1.27 7 1.27 8 1.33 9 1.35 10 1.40 -------- --------- Dynamic range explored: from 993 to 167086 e- Average count rate measured: 355 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.76442E+00 +/- 2.64668E-03 e-/ADU RON = 5.12005E+00 +/- 4.90197E-03 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 1.00000E-02 sec *** MIDAS logfile (vers. 02FEBpl1.0) *** Mo, 08 Sep 2003 17:47:44 Page 3 LINEARITY ERROR =============== Average linearity error after shutter correct. = + 1.0017/ 0.9961 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay