*** MIDAS logfile (vers. 02FEBpl1.0) *** We, 06 Aug 2003 22:37:32 Page 1 *** MIDAS logfile (vers. 02FEBpl1.0) *** We, 06 Aug 2003 22:37:54 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#60 TAKEN ON 20030805 IN BLOW MODE ========================================================== BIAS ==== 5 biases at start test: average = 384.8041 stddev = 1.8620 There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_BLOW_60_20030805 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 1.000 342.7770 342.2480 144.1470 1.0040 1.0026 1.19 2 6.799 2292.0061 2292.8059 948.5850 0.9958 0.9963 1.21 3 15.300 5164.5962 5164.5859 2177.2749 0.9980 0.9981 1.19 4 27.199 9185.2764 9185.9463 3652.0500 0.9987 0.9989 1.26 5 42.500 14366.095 14365.496 5807.1001 0.9998 0.9999 1.24 6 61.200 20700.996 20700.896 8136.7500 1.0005 1.0006 1.27 7 83.299 28187.695 28188.595 10893.450 1.0010 1.0011 1.29 8 108.799 36823.996 36822.996 13930.099 1.0012 1.0013 1.32 9 137.699 46592.394 46593.097 17002.050 1.0009 1.0011 1.37 10 170.000 57468.996 57468.996 20760.800 1.0000 1.0002 1.38 -------- --------- --------- --------- --------- --------- --------- --------- Sequence el_adu2 -------- --------- 1 1.19 2 1.21 3 1.19 4 1.26 5 1.24 6 1.27 7 1.29 8 1.32 9 1.37 10 1.38 -------- --------- Dynamic range explored: from 948 to 159136 e- Average count rate measured: 338 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.76908E+00 +/- 3.40278E-03 e-/ADU RON = 5.15613E+00 +/- 6.33612E-03 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 1.00000E-02 sec *** MIDAS logfile (vers. 02FEBpl1.0) *** We, 06 Aug 2003 22:37:54 Page 3 LINEARITY ERROR =============== Average linearity error after shutter correct. = + 1.0033/ 0.9961 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay