*** MIDAS logfile (vers. 02FEBpl1.0) *** Fr, 25 Jul 2003 16:52:30 Page 1 *** MIDAS logfile (vers. 02FEBpl1.0) *** Fr, 25 Jul 2003 16:52:57 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#60 TAKEN ON 20030724 IN BLOW MODE ========================================================== BIAS ==== 5 biases at start test: average = 378.7232 stddev = 1.9437 There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_BLOW_60_20030724 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 1.000 348.1780 348.0460 143.6495 1.0039 0.9951 1.21 2 6.799 2328.6069 2328.7971 947.3600 0.9960 0.9959 1.23 3 15.300 5246.1768 5245.6470 2105.1101 0.9979 0.9985 1.25 4 27.199 9330.1973 9329.9873 3767.1001 0.9986 0.9996 1.24 5 42.500 14592.676 14592.777 5809.0498 0.9997 1.0008 1.26 6 61.200 21027.777 21027.277 8192.4004 1.0005 1.0016 1.28 7 83.299 28637.976 28634.377 10825.799 1.0011 1.0022 1.32 8 108.799 37409.078 37409.078 14256.250 1.0012 1.0025 1.31 9 137.699 47334.878 47336.277 17288.900 1.0010 1.0024 1.37 10 170.000 58377.578 58375.976 20505.150 1.0000 1.0013 1.42 -------- --------- --------- --------- --------- --------- --------- --------- Sequence el_adu2 -------- --------- 1 1.21 2 1.23 3 1.25 4 1.24 5 1.26 6 1.28 7 1.32 8 1.31 9 1.37 10 1.42 -------- --------- Dynamic range explored: from 978 to 163948 e- Average count rate measured: 343 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.80844E+00 +/- 2.58858E-03 e-/ADU RON = 5.45884E+00 +/- 5.03149E-03 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 1.50000E-02 sec *** MIDAS logfile (vers. 02FEBpl1.0) *** Fr, 25 Jul 2003 16:52:57 Page 3 LINEARITY ERROR =============== Average linearity error after shutter correct. = + 1.0032/ 0.9955 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay