*** MIDAS logfile (vers. 02FEBpl1.1) *** We, 03 Sep 2003 17:29:39 Page 1 *** MIDAS logfile (vers. 02FEBpl1.1) *** We, 03 Sep 2003 17:29:48 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#60 TAKEN ON 20030111 IN AHIGH MODE ========================================================== BIAS ==== 5 biases at start test: average = 418.4079 stddev = 6.2510 There were 186 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_AHIGH_60_20030111 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 0.400 705.5420 707.9220 1354.5150 1.0024 1.0054 0.26 2 1.500 2579.7720 2582.0520 4798.7202 0.9952 0.9957 0.27 3 3.399 5853.5620 5853.8521 10811.099 1.0003 0.9999 0.27 4 6.099 10486.792 10487.891 19291.599 1.0000 0.9997 0.27 5 9.500 16337.292 16337.192 29988.449 1.0007 1.0003 0.27 6 13.699 23555.492 23554.392 43450.101 1.0009 1.0005 0.27 7 18.600 31980.492 31980.191 58734.000 1.0010 1.0006 0.27 8 24.300 41798.293 41793.691 76444.500 1.0016 1.0011 0.27 9 30.800 52976.492 52972.293 95796.500 1.0016 1.0011 0.28 10 38.000 65003.992 65003.093 27243.550 0.9962 0.9958 1.19 -------- --------- --------- --------- --------- --------- --------- --------- Sequence el_adu2 -------- --------- 1 0.26 2 0.27 3 0.27 4 0.27 5 0.27 6 0.27 7 0.27 8 0.27 9 0.28 10 1.19 -------- --------- Dynamic range explored: from 706 to 64915 e- Average count rate measured: 1717 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 9.98641E-01 +/- 4.21404E-05 e-/ADU RON = 6.24255E+00 +/- 2.63421E-04 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 1.00000E-02 sec *** MIDAS logfile (vers. 02FEBpl1.1) *** We, 03 Sep 2003 17:29:48 Page 3 LINEARITY ERROR =============== Average linearity error after shutter correct. = + 1.0039/ 0.9955 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay