*** MIDAS logfile (vers. 02FEBpl1.1) *** We, 03 Sep 2003 19:56:35 Page 1 *** MIDAS logfile (vers. 02FEBpl1.1) *** We, 03 Sep 2003 19:56:48 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#60 TAKEN ON 20030204 IN AABHIGH MODE ========================================================== BIAS ==== 5 biases at start test: average = 525.0220 stddev = 15.5708 There were 186 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_AABHIGH_60_20030204 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 el_adu2 -------- --------- --------- --------- --------- --------- --------- --------- --------- 1 1.000 1733.7679 1743.9780 3421.8149 0.9992 1.0039 0.25 0.25 2 2.000 3423.6580 3443.9180 6555.2500 0.9915 0.9962 0.26 0.26 3 3.399 5852.3179 5856.1182 11062.599 0.9993 0.9988 0.26 0.26 4 6.099 10497.877 10499.177 19687.500 1.0003 0.9992 0.27 0.27 5 9.500 16357.177 16364.278 30389.199 1.0012 1.0005 0.27 0.27 6 13.699 23597.578 23591.078 43505.101 1.0020 1.0005 0.27 0.27 7 18.600 32028.078 32028.978 58687.500 1.0018 1.0007 0.27 0.27 8 24.300 41850.179 41831.578 76285.500 1.0021 1.0005 0.27 0.27 9 30.800 53000.878 52983.078 95100.000 1.0013 0.9998 0.28 0.28 10 35.000 60231.878 60206.476 108159.50 1.0014 0.9999 0.28 0.28 -------- --------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 971 to 33631 e- Average count rate measured: 1719 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 5.58475E-01 +/- 2.04310E-03 e-/ADU RON = 8.69587E+00 +/- 3.18126E-02 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 1.00000E-02 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 1.0030/ 0.9938 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay