*** MIDAS logfile (vers. 02FEBpl1.0) *** We, 03 Sep 2003 20:25:36 Page 1 *** MIDAS logfile (vers. 02FEBpl1.0) *** We, 03 Sep 2003 20:26:17 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#60 TAKEN ON 20030902 IN BLOW MODE ========================================================== BIAS ==== 5 biases at start test: average = 383.1365 stddev = 1.8392 There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_BLOW_60_20030902 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 1.000 358.2110 357.7980 149.3005 1.0019 1.0009 1.20 2 6.799 2400.3630 2400.4431 988.4200 0.9959 0.9961 1.21 3 15.300 5410.2432 5409.3232 2197.6350 0.9983 0.9983 1.23 4 27.199 9621.7627 9621.6631 3879.2351 0.9990 0.9991 1.24 5 42.500 15047.762 15047.863 6069.4502 1.0000 1.0002 1.24 6 61.200 21684.863 21684.562 8945.6504 1.0008 1.0009 1.21 7 83.299 29529.662 29529.363 11220.900 1.0014 1.0015 1.32 8 108.799 38572.363 38573.363 14692.150 1.0015 1.0016 1.31 9 137.699 48800.964 48801.363 17898.599 1.0011 1.0013 1.36 10 170.000 60177.664 60179.062 21599.349 1.0000 1.0001 1.39 -------- --------- --------- --------- --------- --------- --------- --------- Sequence el_adu2 -------- --------- 1 1.20 2 1.21 3 1.23 4 1.24 5 1.24 6 1.21 7 1.32 8 1.31 9 1.36 10 1.39 -------- --------- Dynamic range explored: from 993 to 166900 e- Average count rate measured: 354 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.77342E+00 +/- 2.51321E-03 e-/ADU RON = 5.10092E+00 +/- 4.62234E-03 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 1.00000E-02 sec *** MIDAS logfile (vers. 02FEBpl1.0) *** We, 03 Sep 2003 20:26:17 Page 3 LINEARITY ERROR =============== Average linearity error after shutter correct. = + 1.0018/ 0.9960 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay