*** MIDAS logfile (vers. 02FEBpl1.0) *** Mo, 30 Jun 2003 23:47:57 Page 1 *** MIDAS logfile (vers. 02FEBpl1.0) *** Mo, 30 Jun 2003 23:48:32 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#60 TAKEN ON 20030630 IN BLOW MODE ========================================================== BIAS ==== 5 biases at start test: average = 380.9367 stddev = 1.9228 There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_BLOW_60_20030630 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 1.000 328.9220 329.0270 136.1050 0.9992 0.9995 1.21 2 6.799 2211.0530 2211.7029 902.6000 0.9963 0.9966 1.22 3 15.300 4981.4131 4979.5332 1997.9399 0.9983 0.9979 1.25 4 27.199 8860.8428 8861.7529 3537.4150 0.9992 0.9993 1.25 5 42.500 13858.462 13853.962 5453.1499 1.0002 0.9999 1.27 6 61.200 19966.562 19968.363 7737.5498 1.0008 1.0009 1.29 7 83.299 27195.363 27190.062 10602.400 1.0016 1.0014 1.28 8 108.799 35522.761 35532.164 13487.049 1.0017 1.0019 1.32 9 137.699 44965.863 44957.464 16589.199 1.0018 1.0017 1.36 10 170.000 55466.562 55453.863 20003.699 1.0010 1.0008 1.39 -------- --------- --------- --------- --------- --------- --------- --------- Sequence el_adu2 -------- --------- 1 1.21 2 1.23 3 1.25 4 1.25 5 1.27 6 1.29 7 1.28 8 1.32 9 1.36 10 1.39 -------- --------- Dynamic range explored: from 905 to 152599 e- Average count rate measured: 326 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.75151E+00 +/- 3.53999E-03 e-/ADU RON = 5.29068E+00 +/- 6.80680E-03 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 1.00000E-02 sec *** MIDAS logfile (vers. 02FEBpl1.0) *** Mo, 30 Jun 2003 23:48:33 Page 3 LINEARITY ERROR =============== Average linearity error after shutter correct. = + 1.0019/ 0.9965 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay