*** MIDAS logfile (vers. 02FEBpl1.1) *** We, 09 Jul 2003 04:13:33 Page 1 *** MIDAS logfile (vers. 02FEBpl1.1) *** We, 09 Jul 2003 04:13:41 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#60 TAKEN ON 20021122 IN AHIGH MODE ========================================================== BIAS ==== 5 biases at start test: average = 425.5049 stddev = 6.1364 There were 186 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_AHIGH_60_20021122 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 0.400 680.8150 682.9450 1300.8400 1.0072 0.9893 0.26 2 1.500 2534.3350 2535.8750 4725.3550 0.9998 0.9974 0.27 3 3.399 5741.3550 5739.9751 10669.200 0.9996 1.0000 0.27 4 6.099 10296.294 10295.294 19093.150 0.9990 1.0009 0.27 5 9.500 16035.195 16035.995 29979.900 0.9989 1.0014 0.27 6 13.699 23125.494 23124.794 42273.199 0.9990 1.0018 0.27 7 18.600 31397.894 31397.595 57831.500 0.9990 1.0020 0.27 8 24.300 41029.394 41030.093 74583.500 0.9992 1.0024 0.28 9 30.800 52001.894 52002.093 94870.500 0.9991 1.0024 0.27 10 38.000 64160.996 64158.796 108855.50 0.9992 1.0025 0.29 -------- --------- --------- --------- --------- --------- --------- --------- Sequence el_adu2 -------- --------- 1 0.26 2 0.27 3 0.27 4 0.27 5 0.27 6 0.27 7 0.27 8 0.28 9 0.27 10 0.29 -------- --------- Dynamic range explored: from 391 to 36763 e- Average count rate measured: 1687 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 5.72984E-01 +/- 4.65036E-04 e-/ADU RON = 3.51605E+00 +/- 2.85364E-03 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 5.00000E-03 sec *** MIDAS logfile (vers. 02FEBpl1.1) *** We, 09 Jul 2003 04:13:41 Page 3 LINEARITY ERROR =============== Average linearity error after shutter correct. = + 1.0048/ 0.9941 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay