*** MIDAS logfile (vers. 02FEBpl1.1) *** Th, 19 Jun 2003 23:57:52 Page 1 *** MIDAS logfile (vers. 02FEBpl1.1) *** Th, 19 Jun 2003 23:58:00 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#60 TAKEN ON 20021119 IN AHIGH MODE ========================================================== BIAS ==== 5 biases at start test: average = 424.8911 stddev = 6.2188 There were 186 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_AHIGH_60_20021119 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 0.400 692.4190 693.7590 1338.0800 1.0037 1.0057 0.26 2 1.500 2529.5791 2525.7690 4697.5049 0.9956 0.9941 0.27 3 3.399 5734.0991 5731.5288 10641.900 0.9997 0.9992 0.27 4 6.099 10267.909 10267.209 19230.099 0.9989 0.9989 0.27 5 9.500 15995.409 15996.808 29618.349 0.9996 0.9997 0.27 6 13.699 23068.808 23068.109 42288.550 1.0001 1.0001 0.27 7 18.600 31320.408 31321.308 57920.000 1.0002 1.0003 0.27 8 24.300 40932.808 40924.808 74836.000 1.0007 1.0005 0.27 9 30.800 51868.808 51866.808 93933.500 1.0005 1.0005 0.28 10 38.000 64015.710 64013.007 110070.50 1.0009 1.0009 0.29 -------- --------- --------- --------- --------- --------- --------- --------- Sequence el_adu2 -------- --------- 1 0.26 2 0.27 3 0.27 4 0.27 5 0.27 6 0.27 7 0.27 8 0.27 9 0.28 10 0.29 -------- --------- Dynamic range explored: from 395 to 36452 e- Average count rate measured: 1683 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 5.69428E-01 +/- 5.74766E-04 e-/ADU RON = 3.54118E+00 +/- 3.57437E-03 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 1.00000E-02 sec *** MIDAS logfile (vers. 02FEBpl1.1) *** Th, 19 Jun 2003 23:58:00 Page 3 LINEARITY ERROR =============== Average linearity error after shutter correct. = + 1.0047/ 0.9949 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay