The FEROS at the MPG/ESO-2.20m CCD Page


CCD#60 in mode Amp B, High gain.

Logo

General Information

The CCD camera is operated with the CCD controller developed by the Astronomical Instrument Centre at the AOC and identical to that in operation at the Danish 1.54-m telescope on La Silla. The controller is capable of delivering the full performance of the CCD in terms of readout noise, CTE, immunity from interference, crosstalk, etc., and provides the image files with the necessary header information. Readout times are of the order of 70 sec through two parallel signal chains.

CCD Characteristics
Op. temperature -1100 to -1200 C
RON 3 e- (for 50 kpix/sec/port)
Dark current 1.5 e-/h/pixel
CTE (par/ser) 0.99999/0.99999
Linearity < 1% deviation up to full well
Full well capacity 150.000 e-

The performance figures given above are based on experience at AOC and are typical measured figures, after optimising operational temperatures, clock waveforms, etc. They apply to CCDs operated at a pixel rate of 50 kpixels/sec and at a nominal temperature of -110°C. ``Full well'' is defined as the exposure limit to which no measurable departure from linearity occurs; the exposure level causing actual blooming is considerably higher.

The camera uses an EEV CCD detector of the design specified for FEROS, i.e., 2048x4096 pixels of 15x15microns, mounted in a continuous-flow cryostat. The CCD can only be mounted with the columns in the main dispersion direction.

The thinned 2kx4k CCD and the standard ESO-VLT continuous-flow cryostat are provided by ESO.

The CCD controller electronics (+ 1 spare) were built by the AOC; also the integration and testing of the complete CCD system were carried out at the AOC.

Note regarding the bias level from Peter Sinclair:

My experience with the Danish controller on DFOSC tells me that a change in 5 adu is rather normal... Actually, the bias level of DFOSC moves around a larger figure... within a single test. I have discussed this with Preben Norregard and he says that this should not be a problem provided the user takes the bias level from the overscan area...

* Derivatives
* Standard CCD Test
* Long CCD Test



Time Evolution

The below plot shows the variation of the bias level, standard light-source (note that often a beta-light is used which, due to radioactive decay, will fade. It is also possible that the phosphor can be poisoned, but our data show no evidence for such an effect), gain, read-out noise, shutter error, and linearity error as a function of time. The values for the bias, gain and read-out noise are plotted along with the associated measurement error. The blue dashed line in the plots shows the mean value of all the points, while the dotted blue line shows the standard deviation of this distribution. The blue and red points in the bottom plot show the maximum deviations in a positive and negative sense wrt the least-squares fit of the linearity data. Plotted vertically on the bottom of the graph is the date (dd/mm) on which the test was made.

[Click here to download ps file]

For printing, the above plot is also available in postscript format . The raw data can also be viewed in a tabular format.




Standard CCD Test

Below we show the most recent test.



*** MIDAS logfile (vers. 02FEBpl1.0) *** Su, 17 Nov 2002   15:19:40   Page 1
 
 
*** MIDAS logfile (vers. 02FEBpl1.0) *** Su, 17 Nov 2002   15:20:18   Page 2
 
     ==========================================================
     SUMMARY STANDARD TEST CCD#60 TAKEN ON 20021117 IN BHIGH MODE
     ==========================================================
 
      BIAS
      ====
 
      5 biases at start test: average = 419.95 stddev =   7.85
 
      There were 186 hot pixels/columns detected
 
 
      TRANSFER FUNCTION AND LINEARITY DATA
      ====================================
 
***Warning*** Command Line truncated to 8 parameters
Check Help  READ/TABLE for more info
  Table : Std_BHIGH_60_20021117 
 Sequence Exp_tim1  Med_cnt1  Med_cnt2  Variance  Frc_cnt1  Frc_cnt2  el_adu1  
 -------- --------- --------- --------- --------- --------- --------- ---------
        1     0.400  637.9420  643.7320 1235.0200    0.9995    1.0077      0.26
        2     1.500 2443.6819 2443.9419 4515.7100    1.0021    1.0013      0.27
        3     3.399 5540.4219 5540.5322 10199.049    0.9989    0.9980      0.27
        4     6.099 9955.5518 9957.8516 18303.349    0.9990    0.9984      0.27
        5     9.500 15520.952 15519.351 28627.599    0.9993    0.9983      0.27
        6    13.699 22392.351 22392.451 40626.250    0.9995    0.9986      0.28
        7    18.600 30420.752 30420.752 55059.500    0.9999    0.9990      0.28
        8    24.300 39758.453 39756.253 71029.000    1.0002    0.9992      0.28
        9    30.800 50419.851 50417.453 89532.000    1.0006    0.9996      0.28
       10    38.000 62227.851 62220.050 111713.50    1.0009    0.9998      0.28
 -------- --------- --------- --------- --------- --------- --------- ---------
 
      Dynamic range explored: from 359 to 34903 e-
      Average count rate measured: 1637 ADU/sec.
 
      CONV. FACTOR AND RON
      ====================
 
      C.Fact =   0.56 +/-   0.00 e-/ADU
 
      RON    =   4.40 +/-   0.00 e- RMS
 
 
      SHUTTER ERROR
      =============
 
 Average shutter error derived from fractional count rate =  0.000 sec
 
 
      LINEARITY ERROR
      ===============
 
 Average linearity error after shutter correct. = + 1.00/ 1.00 %
 
 These errors represent maximum deviation in % from the average
 count rate after correction for shutter delay

In addition, the previous standard tests are available. Click in the table below to access these files.

Standard Tests 1 2 3 4 5 6
20021117.txt aver.ps cbias.ps hist.ps lin1.ps lin2.ps trans.ps

In the above table, the postscript files are:

Column
Number
Description
1 Average of the rows and cols. of the bias image
2 Combined bias frame of 5 bias images
3 Histogram of the bias image
4 Plots of exposure sequence, unedited transfer and raw linearity
5 Plots of filtered linearity and differential non-linearity in
6 Filtered transfer curves for both data sets

Note: filtered means that saturated images or deviating too much from the other data, if present, have been removed.




Long CCD Test

Below are the results of the long tests carried out by ODT. Be warned that these are large compressed tar files, and are of specialised interest. To download, press the shift key and click on the hypertext link with your middle mouse button.

Long Tests


Updated:
17/Nov/2002
2p2 Team
CCD Page
2p2 Team
Home
Comments to:
2p2team@eso.org