The FEROS at the MPG/ESO-2.20m CCD Page


CCD#60 in mode Amp A, Low gain.

Logo

General Information

The CCD camera is operated with the CCD controller developed by the Astronomical Instrument Centre at the AOC and identical to that in operation at the Danish 1.54-m telescope on La Silla. The controller is capable of delivering the full performance of the CCD in terms of readout noise, CTE, immunity from interference, crosstalk, etc., and provides the image files with the necessary header information. Readout times are of the order of 70 sec through two parallel signal chains.

CCD Characteristics
Op. temperature -1100 to -1200 C
RON 3 e- (for 50 kpix/sec/port)
Dark current 1.5 e-/h/pixel
CTE (par/ser) 0.99999/0.99999
Linearity < 1% deviation up to full well
Full well capacity 150.000 e-

The performance figures given above are based on experience at AOC and are typical measured figures, after optimising operational temperatures, clock waveforms, etc. They apply to CCDs operated at a pixel rate of 50 kpixels/sec and at a nominal temperature of -110°C. ``Full well'' is defined as the exposure limit to which no measurable departure from linearity occurs; the exposure level causing actual blooming is considerably higher.

The camera uses an EEV CCD detector of the design specified for FEROS, i.e., 2048x4096 pixels of 15x15microns, mounted in a continuous-flow cryostat. The CCD can only be mounted with the columns in the main dispersion direction.

The thinned 2kx4k CCD and the standard ESO-VLT continuous-flow cryostat are provided by ESO.

The CCD controller electronics (+ 1 spare) were built by the AOC; also the integration and testing of the complete CCD system were carried out at the AOC.

Note regarding the bias level from Peter Sinclair:

My experience with the Danish controller on DFOSC tells me that a change in 5 adu is rather normal... Actually, the bias level of DFOSC moves around a larger figure... within a single test. I have discussed this with Preben Norregard and he says that this should not be a problem provided the user takes the bias level from the overscan area...

* Derivatives
* Standard CCD Test
* Long CCD Test



Time Evolution

The below plot shows the variation of the bias level, standard light-source (note that often a beta-light is used which, due to radioactive decay, will fade. It is also possible that the phosphor can be poisoned, but our data show no evidence for such an effect), gain, read-out noise, shutter error, and linearity error as a function of time. The values for the bias, gain and read-out noise are plotted along with the associated measurement error. The blue dashed line in the plots shows the mean value of all the points, while the dotted blue line shows the standard deviation of this distribution. The blue and red points in the bottom plot show the maximum deviations in a positive and negative sense wrt the least-squares fit of the linearity data. Plotted vertically on the bottom of the graph is the date (dd/mm) on which the test was made.

[Click here to download ps file]

For printing, the above plot is also available in postscript format . The raw data can also be viewed in a tabular format.




Standard CCD Test

Below we show the most recent test.



*** MIDAS logfile (vers. 02FEBpl1.0) *** Th, 14 Nov 2002   18:56:26   Page 1
 
 
*** MIDAS logfile (vers. 02FEBpl1.0) *** Th, 14 Nov 2002   18:57:06   Page 2
 
     ==========================================================
     SUMMARY STANDARD TEST CCD#60 TAKEN ON 20021114 IN ALOW MODE
     ==========================================================
 
      BIAS
      ====
 
      5 biases at start test: average = 384.06 stddev =   2.52
 
      There were 11 hot pixels/columns detected
 
 
      TRANSFER FUNCTION AND LINEARITY DATA
      ====================================
 
***Warning*** Command Line truncated to 8 parameters
Check Help  READ/TABLE for more info
  Table : Std_ALOW_60_20021114 
 Sequence Exp_tim1  Med_cnt1  Med_cnt2  Variance  Frc_cnt1  Frc_cnt2  el_adu1  
 -------- --------- --------- --------- --------- --------- --------- ---------
        1     2.000  740.0830  740.2430  307.7300    1.0001    1.0003      1.20
        2     4.000 1477.9530 1477.8030  601.0750    0.9986    0.9985      1.23
        3     8.000 2959.0830 2958.9729 1197.6200    0.9997    0.9997      1.24
        4    16.000 5920.1230 5919.6831 2434.9700    1.0000    1.0000      1.22
        5    32.000 11846.342 11844.842 4691.9751    1.0005    1.0004      1.26
        6    54.000 19996.043 19996.843 7807.7002    1.0008    1.0009      1.28
        7    78.000 28885.642 28885.142 11234.950    1.0009    1.0009      1.29
        8   102.000 37759.742 37760.941 14330.150    1.0005    1.0006      1.32
        9   128.000 47356.543 47356.144 17808.750    0.9999    0.9999      1.33
       10   156.000 57649.644 57650.043 20558.900    0.9988    0.9988      1.40
 -------- --------- --------- --------- --------- --------- --------- ---------
 
      Dynamic range explored: from 2032 to 158240 e-
      Average count rate measured: 370 ADU/sec.
 
      CONV. FACTOR AND RON
      ====================
 
      C.Fact =   2.74 +/-   0.00 e-/ADU
 
      RON    =   6.91 +/-   0.00 e- RMS
 
 
      SHUTTER ERROR
      =============
 
 Average shutter error derived from fractional count rate =  0.000 sec
 
 
      LINEARITY ERROR
      ===============
 
 Average linearity error after shutter correct. = + 1.00/ 1.00 %
 
 These errors represent maximum deviation in % from the average
 count rate after correction for shutter delay

In addition, the previous standard tests are available. Click in the table below to access these files.

Standard Tests 1 2 3 4 5 6
20021023.txt aver.ps cbias.ps hist.ps lin1.ps lin2.ps trans.ps
20021114.txt aver.ps cbias.ps hist.ps lin1.ps lin2.ps trans.ps

In the above table, the postscript files are:

Column
Number
Description
1 Average of the rows and cols. of the bias image
2 Combined bias frame of 5 bias images
3 Histogram of the bias image
4 Plots of exposure sequence, unedited transfer and raw linearity
5 Plots of filtered linearity and differential non-linearity in
6 Filtered transfer curves for both data sets

Note: filtered means that saturated images or deviating too much from the other data, if present, have been removed.




Long CCD Test

Below are the results of the long tests carried out by ODT. Be warned that these are large compressed tar files, and are of specialised interest. To download, press the shift key and click on the hypertext link with your middle mouse button.

Long Tests


Updated:
22/Nov/2002