| The FEROS at the MPG/ESO-2.20m CCD PageCCD#60 in mode Amp A, High gain. |
|
The CCD camera is operated with the CCD controller developed by the Astronomical Instrument Centre at the AOC and identical to that in operation at the Danish 1.54-m telescope on La Silla. The controller is capable of delivering the full performance of the CCD in terms of readout noise, CTE, immunity from interference, crosstalk, etc., and provides the image files with the necessary header information. Readout times are of the order of 70 sec through two parallel signal chains.
| CCD Characteristics | |
|---|---|
| Op. temperature | -1100 to -1200 C |
| RON | 3 e- (for 50 kpix/sec/port) |
| Dark current | 1.5 e-/h/pixel |
| CTE (par/ser) | 0.99999/0.99999 |
| Linearity | < 1% deviation up to full well |
| Full well capacity | 150.000 e- |
The performance figures given above are based on experience at AOC and are typical measured figures, after optimising operational temperatures, clock waveforms, etc. They apply to CCDs operated at a pixel rate of 50 kpixels/sec and at a nominal temperature of -110°C. ``Full well'' is defined as the exposure limit to which no measurable departure from linearity occurs; the exposure level causing actual blooming is considerably higher.
The camera uses an EEV CCD detector of the design specified for FEROS, i.e., 2048x4096 pixels of 15x15microns, mounted in a continuous-flow cryostat. The CCD can only be mounted with the columns in the main dispersion direction.
The thinned 2kx4k CCD and the standard ESO-VLT continuous-flow cryostat are provided by ESO.
The CCD controller electronics (+ 1 spare) were built by the AOC; also the integration and testing of the complete CCD system were carried out at the AOC.
Note regarding the bias level from Peter Sinclair:
My experience with the Danish controller on DFOSC tells me that a change
in 5 adu is rather normal... Actually, the bias level of DFOSC moves
around a larger figure... within a single test. I have discussed this
with Preben Norregard and he says that this should not be a problem
provided the user takes the bias level from the overscan area...
Derivatives
Standard CCD Test
Long CCD Test
The below plot shows the variation of the bias level, standard light-source (note that often a beta-light is used which, due to radioactive decay, will fade. It is also possible that the phosphor can be poisoned, but our data show no evidence for such an effect), gain, read-out noise, shutter error, and linearity error as a function of time. The values for the bias, gain and read-out noise are plotted along with the associated measurement error. The blue dashed line in the plots shows the mean value of all the points, while the dotted blue line shows the standard deviation of this distribution. The blue and red points in the bottom plot show the maximum deviations in a positive and negative sense wrt the least-squares fit of the linearity data. Plotted vertically on the bottom of the graph is the date (dd/mm) on which the test was made.
For printing, the above plot is also available in postscript format . The raw data can also be viewed in a tabular format.
Below we show the most recent test.
*** MIDAS logfile (vers. 02FEBpl1.0) *** Su, 17 Nov 2002 15:20:43 Page 1
*** MIDAS logfile (vers. 02FEBpl1.0) *** Su, 17 Nov 2002 15:21:21 Page 2
==========================================================
SUMMARY STANDARD TEST CCD#60 TAKEN ON 20021117 IN AHIGH MODE
==========================================================
BIAS
====
5 biases at start test: average = 425.09 stddev = 6.20
There were 186 hot pixels/columns detected
TRANSFER FUNCTION AND LINEARITY DATA
====================================
***Warning*** Command Line truncated to 8 parameters
Check Help READ/TABLE for more info
Table : Std_AHIGH_60_20021117
Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1
-------- --------- --------- --------- --------- --------- --------- ---------
1 0.400 674.8990 675.2590 1298.4449 1.0039 1.0044 0.26
2 1.500 2514.7090 2514.8191 4660.3501 0.9975 0.9975 0.27
3 3.399 5707.2690 5707.8188 10494.700 0.9990 0.9991 0.27
4 6.099 10241.308 10241.709 19012.050 0.9991 0.9991 0.27
5 9.500 15960.808 15961.308 29592.849 0.9996 0.9996 0.27
6 13.699 23019.209 23020.808 42204.500 0.9998 0.9998 0.27
7 18.600 31261.509 31258.709 57098.000 1.0000 0.9999 0.27
8 24.300 40851.808 40850.210 72977.500 1.0003 1.0002 0.28
9 30.800 51779.910 51781.808 92944.000 1.0003 1.0002 0.28
10 38.000 63902.910 63895.308 112466.50 1.0006 1.0004 0.28
-------- --------- --------- --------- --------- --------- --------- ---------
Dynamic range explored: from 382 to 36140 e-
Average count rate measured: 1681 ADU/sec.
CONV. FACTOR AND RON
====================
C.Fact = 0.57 +/- 0.00 e-/ADU
RON = 3.51 +/- 0.00 e- RMS
SHUTTER ERROR
=============
Average shutter error derived from fractional count rate = 0.000 sec
LINEARITY ERROR
===============
Average linearity error after shutter correct. = + 1.00/ 1.00 %
These errors represent maximum deviation in % from the average
count rate after correction for shutter delay
In addition, the previous standard tests are available. Click in the table below to access these files.
| Standard Tests | 1 | 2 | 3 | 4 | 5 | 6 |
|---|---|---|---|---|---|---|
| 20021117.txt | aver.ps | cbias.ps | hist.ps | lin1.ps | lin2.ps | trans.ps |
In the above table, the postscript files are:
| Column Number | Description |
|---|---|
| 1 | Average of the rows and cols. of the bias image |
| 2 | Combined bias frame of 5 bias images |
| 3 | Histogram of the bias image |
| 4 | Plots of exposure sequence, unedited transfer and raw linearity |
| 5 | Plots of filtered linearity and differential non-linearity in |
| 6 | Filtered transfer curves for both data sets |
Note: filtered means that saturated images or deviating too much from the other data, if present, have been removed.
Below are the results of the long tests carried out by ODT. Be warned that these are large compressed tar files, and are of specialised interest. To download, press the shift key and click on the hypertext link with your middle mouse button.
| Long Tests |
|---|
| Updated: 17/Nov/2002 |