| Boller & Chivens CCD Page 2 |
|
Last UV flooding has been made in date February 23th, 1998 by the
ODT on La Silla.
The last short test on the CCD (June 13th, 1998)
show no particular change in the CCD sensitivity since the last UV
flooding. One of the (undesired) effects of UV flooding is the
broadening of the spectral lines and, therefore, of the instrumental
line width. The amplitude of this effect decreases with wavelength.
The following plot gives an idea of the relationship between the real
slit width and the instrumental FWHM. The measurements were made with
grating #26 (grating angle 19o) and using 8 He-Ar lines form
492.1nm to 573.9nm. Dots represents the FHWM of the 8 lines, as
measured by mean of the ALICE MIDAS context, as a function of slit width
in pixels. The line represents an empirical fit to the data. The
empirical relation is shown at the top of the plot.
Description
Plots and Images
Raw bias frame (fits file or gif file).
Average bias frame (180 frames) (fits file or gif file). Note that on this
frame the rms around the bias value is of the
order of 0.32 ADU and therefore the vertical structures
that is possible to see here are much below the R.O.N of
raw frames.
Bias profile along the column direction. By averaging all
columns of the average bias frame (see previous blue
point) we obtain a high S/N column profile. A
gradient is present toward the upper part of the frame
(i.e. toward the red) of about 0.0006 ADU/pixel. Also
a faint periodic signal is present. The power spectrum of the
Fourier transform indicates that the spatial period of
this signal is of 29.41 pixels.
Bias profile along the row direction. By averaging all rows
of the average bias frame we obtain a high S/N column profile. In
the first 20 columns a steep gradient is present.
Looking the profile in more detail, we can
also notice the gradient of about 0.0011 ADU/pixels and
the narrows peaks giving rise to the vertical structure
visible in the average bias
frame
Bias level as function of time on
raw frames and average frames. The
mean value as been computed in a region of 95x70 pixels
in the center of the CCD. Each average frames has been
obtained combining 10 raw bias frames (a
-
clipping algorithm has been used to avoid cosmic rays).
Read Out Noise as
function of time. The same area 95x70 pixels has been
used to compute this value on raw frames.
|
ESO 1.5-m Home Page |
2p2 Team CCD Page |
2p2 Team Home Page | Last update: Oct.98. Comments to 2p2team@eso.org |
|---|