*** MIDAS logfile (vers. 97NOV) *** Sa, 24 Aug 2002 13:05:02 Page 1 *** MIDAS logfile (vers. 97NOV) *** Sa, 24 Aug 2002 13:06:15 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#38 TAKEN ON 020824 IN slow MODE ========================================================== BIAS ==== 5 biases at start test: average = 213.10 stddev = 4.98 There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_slow_38_020824 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 1.00 919.0 920.2 636.7 -0.06 -0.05 1.44 2 4.00 3672.0 3671.9 2469.4 0.14 0.09 1.49 3 8.01 7344.7 7344.6 4819.0 0.07 0.16 1.52 4 12.00 11000.1 11003.1 7273.9 0.06 0.05 1.51 5 18.00 16504.2 16506.8 10853.4 0.10 0.08 1.52 6 24.00 21991.0 21995.7 14357.2 0.04 0.03 1.53 7 30.00 27476.0 27480.7 17822.2 -0.01 -0.02 1.54 8 41.00 37540.7 37548.3 24209.5 -0.03 -0.04 1.55 9 50.00 45718.9 45729.9 29227.8 -0.16 -0.17 1.56 10 60.00 54873.1 54895.3 34850.5 -0.14 -0.13 1.57 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 1448 to 86452 e- Average count rate measured: 916 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 1.58 +/- 0.01 e-/ADU RON = 7.84 +/- 0.03 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.004 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.15/-0.17 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay