*** MIDAS logfile (vers. 97NOV) *** Tu, 12 Jun 2001 16:25:39 Page 1 *** MIDAS logfile (vers. 97NOV) *** Tu, 12 Jun 2001 16:26:41 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#38 TAKEN ON 010612 IN slow MODE ========================================================== BIAS ==== 5 biases at start test: average = 212.66 stddev = 4.52 There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_slow_38_010612 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 2.30 2035.5 2027.8 1367.3 -0.03 -0.07 1.49 2 4.61 4062.9 4061.2 2686.0 -0.14 -0.02 1.51 3 9.20 8129.5 8118.5 5323.0 0.28 0.22 1.53 4 13.80 12174.6 12173.1 7958.2 0.16 0.13 1.53 5 20.70 18258.2 18254.7 11850.9 0.18 0.14 1.54 6 27.61 24324.2 24321.9 15730.7 0.08 0.12 1.55 7 34.51 30402.6 30392.4 19557.8 0.09 0.10 1.55 8 46.01 40468.2 40466.6 25860.8 -0.07 -0.06 1.56 9 57.50 50603.2 50614.0 32457.2 0.00 0.01 1.56 10 69.01 60401.6 60370.1 4152.0 -0.54 -0.58 14.55 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 6700 to 199137 e- Average count rate measured: 880 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 3.30 +/- 1.64 e-/ADU RON = 14.92 +/- 7.43 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.010 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.25/-0.56 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay