*** MIDAS logfile (vers. 97NOV) *** Su, 08 Oct 2000 18:53:25 Page 1 *** MIDAS logfile (vers. 97NOV) *** Su, 08 Oct 2000 18:54:24 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#38 TAKEN ON 001008 IN slow MODE ========================================================== BIAS ==== 5 biases at start test: average = 213.60 stddev = 4.50 There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_slow_38_001008 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 2.25 1985.2 1985.8 1327.7 -0.14 -0.10 1.50 2 4.48 3962.3 3961.2 2663.9 0.17 0.15 1.49 3 8.96 7920.2 7926.3 5189.5 0.15 0.12 1.53 4 13.44 11874.9 11873.9 7817.4 0.12 0.11 1.52 5 20.16 17811.1 17809.0 11439.4 0.12 0.11 1.56 6 26.89 23729.9 23728.6 15258.2 0.00 0.04 1.56 7 33.60 29655.0 29662.0 19228.1 0.02 0.02 1.54 8 44.80 39491.4 39490.1 25228.7 -0.10 -0.10 1.57 9 56.01 49396.8 49386.8 31920.3 -0.05 -0.05 1.55 10 67.20 59127.2 59129.8 37552.6 -0.28 -0.28 1.57 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3117 to 92834 e- Average count rate measured: 882 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 1.57 +/- 0.01 e-/ADU RON = 7.06 +/- 0.03 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.003 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.16/-0.28 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay