*** MIDAS logfile (vers. 97NOV) *** Sa, 15 Jul 2000 17:57:09 Page 1 *** MIDAS logfile (vers. 97NOV) *** Sa, 15 Jul 2000 17:58:29 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#38 TAKEN ON 000715 IN slow MODE ========================================================== BIAS ==== 5 biases at start test: average = 213.81 stddev = 4.83 There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_slow_38_000715 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 2.16 1987.4 1988.7 1338.0 -0.11 0.01 1.49 2 4.32 3972.3 3970.6 2670.1 0.06 -0.18 1.49 3 8.64 7945.3 7943.6 5231.5 0.19 0.19 1.52 4 12.96 11901.1 11900.2 7921.9 0.08 0.09 1.50 5 19.44 17850.2 17847.1 11771.5 0.10 0.10 1.52 6 25.92 23784.0 23783.3 15438.1 0.04 0.06 1.54 7 32.40 29728.0 29729.9 19079.0 0.04 0.07 1.56 8 43.20 39580.7 39582.8 25349.7 -0.09 -0.07 1.56 9 54.00 49489.0 49502.8 31294.4 -0.06 -0.02 1.58 10 64.80 59268.7 59272.7 37294.3 -0.26 -0.25 1.59 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3169 to 94475 e- Average count rate measured: 917 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 1.59 +/- 0.01 e-/ADU RON = 7.70 +/- 0.04 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.009 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.19/-0.25 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay