*** MIDAS logfile (vers. 97NOV) *** Th, 06 Jul 2000 17:05:02 Page 1 *** MIDAS logfile (vers. 97NOV) *** Th, 06 Jul 2000 17:06:20 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#38 TAKEN ON 000707 IN slow MODE ========================================================== BIAS ==== 5 biases at start test: average = 213.45 stddev = 4.73 There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_slow_38_000707 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 2.21 1997.1 1989.3 1362.8 -0.18 -0.12 1.47 2 4.40 3989.6 3988.3 2680.3 0.20 0.17 1.49 3 8.80 7979.3 7977.9 5287.5 0.23 0.21 1.51 4 13.20 11947.8 11946.9 7978.5 0.06 0.05 1.50 5 19.80 17929.2 17928.1 11536.6 0.10 0.10 1.55 6 26.40 23889.6 23889.6 15635.2 0.04 0.04 1.53 7 33.00 29857.3 29851.1 19208.8 0.03 0.00 1.55 8 44.00 39757.1 39760.1 25493.4 -0.11 -0.10 1.56 9 55.00 49702.3 49715.2 31544.8 -0.09 -0.07 1.58 10 66.00 59533.1 59539.7 37551.2 -0.28 -0.27 1.59 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3167 to 94588 e- Average count rate measured: 904 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 1.59 +/- 0.01 e-/ADU RON = 7.51 +/- 0.04 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.002 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.22/-0.27 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay