2p2-m Telescope CCDs
|
|||||
|
The CCD currently in use with DFOSC at the Danish 1.54-m Telescope was provided by the Copenhagen University Observatory and is a 2kx4k ????.
Observations with CCD02 began on September 20 2000, prior to which CCD01 was in use.
The below plot shows the variation of the bias level, standard light-source (note that often a beta-light is used which, due to radioactive decay, will fade. It is also possible that the phosphor can be poisoned, but our data show no evidence for such an effect), gain, read-out noise, shutter error, and linearity error as a function of time. The values for the bias, gain and read-out noise are plotted along with the associated measurement error. The blue dashed line in the plots shows the mean value of all the points, while the dotted blue line shows the standard deviation of this distribution. The blue and red points in the bottom plot show the maximum deviations in a positive and negative sense wrt the least-squares fit of the linearity data. Plotted vertically on the bottom of the graph is the date (dd/mm) on which the test was made.
For printing, the above plot is also available in postscript format . The raw data can also be viewed in a tabular format.
Below we show the most recent test.
*** MIDAS logfile (vers. 95NOV) *** We, 23 Oct 2002 14:43:53 Page 1 *** MIDAS logfile (vers. 95NOV) *** We, 23 Oct 2002 14:45:25 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#02 TAKEN ON 021023 IN fast MODE ========================================================== BIAS ==== 5 biases at start test: average = 289.65 stddev = 4.21 There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_fast_02_021023 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 1.00 1776.9 1776.7 2419.7 0.02 0.01 0.73 2 2.00 3534.0 3534.3 4830.8 0.00 0.02 0.73 3 4.00 7048.6 7047.8 9616.8 0.00 -0.01 0.73 4 6.00 10565.2 10563.9 14284.9 0.02 0.01 0.74 5 9.00 15834.3 15833.5 21355.7 0.00 -0.01 0.74 6 12.00 21110.9 21108.9 28438.3 0.02 0.01 0.74 7 15.00 26379.7 26379.9 35743.5 0.01 0.01 0.74 8 20.00 35163.2 35167.1 46923.2 0.00 0.01 0.75 9 25.00 43936.0 43939.2 58218.5 -0.03 -0.02 0.75 10 30.00 52719.5 52719.1 68791.1 -0.03 -0.03 0.77 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 1355 to 40214 e- Average count rate measured: 1757 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 0.76 +/- 0.00 e-/ADU RON = 3.21 +/- 0.02 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.011 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.02 /-0.03 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay
In addition, the previous standard tests are available. Click in the table below to access these files.
In the above table, the postscript files are:
Column Number | Description |
---|---|
1 | Average of the rows and cols. of the bias image |
2 | Combined bias frame of 5 bias images |
3 | Histogram of the bias image |
4 | Plots of exposure sequence, unedited transfer and raw linearity |
5 | Plots of filtered linearity and differential non-linearity in |
6 | Filtered transfer curves for both data sets |
Note: filtered means that saturated images or deviating too much from the other data, if present, have been removed.
Below are the results of the long tests carried out by ODT. Be warned that these are large compressed tar files, and are of specialised interest. To download, press the shift key and click on the hypertext link with your middle mouse button.
Long Tests |
---|
|