 |
EFOSC2
ESO
Faint Object Spectrograph and Camera
|
Simple
Imaging
Jittered
Imaging
Coronographic
Imaging
Polarimetric
Imaging
The 4 science imaging modes are described below. The observer is also
referred to the pages on Acquisition images and
Miscellaneous information.
In this mode one can obtain one or more images of a field through broad
and/or narrow band filters.
The following acquisition templates may be used in this mode:
-
Simple Preset
-
Move To Pixel
A sample jP2PP
observing block
The above example shows an OB comprising a sequence of:
-
A simple telescope preset to target with Rotator offset angle = 0 degree
-
1 R-band image of 300 seconds with normal
readout and 2x2 binning followed by
-
3 U-band image of 1200 seconds with normal
readout and 2x2 binning
The FITS file is called EFOSC_Image.#.fits
Parameter Values One does not have to (in fact, should
not) change the value of any parameter other than :
-
Filter: See the Filter
Set for a list of the available choices
-
Exposure Time: This is a list of exposure
times. Whatever appropriate in seconds - keep in mind the reflections of
heavily saturated stars and cosmic ray hits.
-
CCD readout speed: normal , fast
or slow .
-
CCD binning: 2x2 or 1x1
-
Number of exposures: whatever appropriate.
If the number of elements in the exposure time list is less than the number
of exposures the procedure cycles through list from the beginning. For
example if the Exposure time entry was 300 600 and the number
of exposures were 4 then the 4 images would be observed with exposure times
of 300, 600, 300 and 600 seconds.
This is very similar to Simple Imaging
except for the folowing difference :
-
A different template is used
-
The telescope is moved by a specified amount between each exposure in the
template. The telescope offsets are cumulative.
A sample jP2PP observing
block (the OB has been erroneously named Coronographic imaging)
The above example shows an OB comprising a sequence of
-
An acquisition template including preset , focus and with the rotator
offset angle = 0 degrees
-
This template will move a selected object or location in the field to the
pixel (1100, 1024)
-
4 B-band images of 900 seconds each taken with
normal
readout and 2x2 binning. The 4 images were taken with cumulative
telescope offsets of (RA,DEC) : (-10, -10) , (20, 0), (0, 20) and
(-20, 0) - i.e. with reference to the original pointing the locations
are (-10, -10), (10, -10), (10, 10) and (-10, 10). The telescope is returned
to the original pointing at the end of the sequence.
The FITS file is called EFOSC_ImaJit.#.fits
Parameter Values In addition to the parameters
listed under Simple Imaging the following may need to be changed :
-
List of Telescope RA Offsets: These are offsets
in RA from one image to the next (and not from the initial pointing). Note
that the first image will not be taken at the initial pointing unless the
first value in the list is set to 0. In case of fewer elements in the list
of RA offsets than the number of exposures then the procedure repeats the
sequence of offsets.
-
List of Telescope DEC offsets: similar to
the previous parameter.
-
Return to Origin: If set to True
the telescope returns to the original pointing - this would be helpful
if one has several Jittered Imaging templates and does not want the field
to fall off the edge of the CCD. If set to False the telescope will
continue from the position of the last exposure of the previous template.
The Coronographic mode is very similar to the Simple
Imaging Mode except for the following differences :
-
The Coronographic imaging templates lacks the Starplate option available
for the Simple Imaging template. This is because the Coronographic procedure
automatically puts in the coronographic mask into the light path.
-
One has to use the Move To Pixel acquisition template since
the object has to be accurately placed behind the coronographic mask. Note
that the mask is not in place during the acquisition image. Obviously,
the pixel specified in the Move To Pixel template should be the centre
of the coronographic mask (see Coronographic Calibrations)
-
Typically Coronographic targets are extremely bright stars and the acquisition
template is taken without the mask - one should use very small integration
times in the acquisition template.
-
The bright targets and the consequent saturation may result in small deviation
from ideal positioning using the template. In such cases one may have to
position the telescope using trial-and-error offsets. We recommend the
following procedure (see the example OB below) :
-
Include in the OB a very short duration Coronographic Imaging template
between the acquisition template and the programme Coronographic imaging
templates.
-
Execute the OB and terminate it after the acquisition template - this will
position the star close to the centre of the coronographic mask
-
Execute the OB again but skip the acquisition template and go directly
to the very short exposure template.
-
If the star looks properly positioned in this short exposure continue with
the programme templates
-
If not, terminate the OB, ask the telescope operator to apply the desired
telescope offset using the telescope control panel and repeat the OB (after
skipping the acquisition)
A sample jP2PP observing
block
-
The above example shows an OB comprising a sequence of:
-
A Move To Pixel Acquisition template with a very short exposure to avoid
saturating the bright coronographic target star and a target location at
pixel (1150, 1050) which was determined to be the centre of the coronographic
mask
-
A short duration coronographic imaging template to carry out fine adjustments
to the target position (see the procedure described above)
-
3 B-band coronographic images of 600 second exposure
time , normal readout and 2x2 binning
followed by
-
1 R-band coronographic images of 600 second exposure
time , normal readout and 2x2 binning
The FITS file is called EFOSC_ImaCor.#.fits
Parameter Values see under simple imaging
The acquisition template to be used is Img_Acq_Pol.
The imaging is similar to Simple Imaging
except for the following additional features :
-
The Starplate is not free but contains a Wollaston Mask
-
The Grism wheel contains the corresponding Wollaston Grism
-
The Half Wave Plate should be moved in (set to True)
-
The continuous rotation of half wave plate should be
-
set to True to get depolarise the radiation. Note:
in this case there has to be 1 and only 1 value (any value is fine)
entered against
List of HWP rotator positions
-
set to False to keep it fixed. In this case there has to be at least
1 value entered against List of HWP rotator positions
Usually the value entered is a list of 4 angles 0 22.5 45.0 67.5.
Note: in this case the template takes one
exposure at each of the 4 HWP positions for each exposure requested.
-
Finally, only half the field is imaged in any exposure. A second exposure
has to be taken if the remaining half also needs to be imaged.
Refer to Polarisation optics.
IMPORTANT When the half wave plate (HWP)
is set to continuous rotation in a template the counter ends up with 360xN
degrees. When a signal is sent to set the HWP to some small fixed value
(say, 45 degrees) it tries to undo all the N rotations and in doing
so triggers a time-out error. So we have included a command at the end
of every Polarimetry template to initialise the HWP if it was set into
continuous rotation mode - this just takes a few seconds. However when
such a template is aborted, for whatever reason, this initialisation does
not take place. The only solution is to cold-start the EFOSC2 instrument
which takes about 1-2 minutes. So avoid aborting a Polarimetry template
- use the STOP option instead.
A sample jP2PP
observing block
The above example shows an OB comprising a sequence of:
-
A Polarimetric imaging acquisition template: the procedure will pop up
a panel at run-time seeking the pixel value where the target has to be
located. This pixel value has to be determined using the procedure described
in Polarimetric calibrations.
-
2 V-band Polarimetric images of 300 second exposure
time with 20" perpendicular Wollaston mask and prism , normal
readout and 2x2 binning at each of 4 half
wave plate positions (at 0, 22.5, 45 and 67.5 degrees) , i.e. 8 exposures
in all , followed by
-
1 V-band Polarimetric image of 300 second exposure
time with 20" perpendicular Wollaston mask and prism, normal
readout and 2x2 binning with the Half Wave Plate
in continuous rotation mode.
back to top