[ ESO ] La Silla

EUROPEAN SOUTHERN OBSERVATORY
SciOp
La Silla - Science Operations Department


FEROS on the MPG/ESO-2.20m

Performance

Quality Control


norm
slow
fast
norm 2x2 binning
slow 2x2 binning
Daily Health Check 225kHz,1,low,1x1 N/A N/A 225kHz,1,low,2x2 N/A
Standard Calibrations 225kHz,1,low,1x1 60kHz,1,high,1x1 625kHz,1,med,1x1 225kHz,1,low,2x2 60kHz,1,high,2x2
Instrument Efficiency 225kHz,1,low,1x1 60kHz,1,high,1x1 625kHz,1,med,1x1 225kHz,1,low,2x2 60kHz,1,high,2x2

CCDTests are here.

Daily Health Check

Since 2005-Nov-19, as part of the Calibration Plan, a Daily Health Check OB consisting of: is executed each afternoon. This OB allows the verification of the health of the instrument and lamps as well as permitting an initialisation of the FEROS DRS for both 1x1 and 2x2 readout modes.

Standard Calibration Sets

As part of the Calibration Plan, sets of Standard Calibration consisting of: are acquired each morning in each readout mode used during the night.

Instrument Efficiency

As part of the Calibration Plan, the instrument efficiency, or more precisely the instrument throughput is measured for every night with FEROS SCIENCE observations via the observation of Spectrophotometric Standard stars.

ADC efficiency

The black curve shows the ADC efficiency. The transmission shows a decrease below 400 nm. Therefore, a new ADC with a better trasnmission in the blue part of the spectrum (red curve) has been orderer and is foreseen to replace the current ADC from period 78 on.
 


[Observing]  [ESO La Silla]  [The SciOps Home Page]  [ESO]  [Index]  [Search]  [Help]  [News]